How to track gage calibration

What goes in the register, how to set intervals, how to label gages, and what to do when one comes back out of tolerance.

Every gage that decides whether a part is good needs its own ID, a known calibration status and a due date, all kept in one register that someone actually checks. Set each interval from the maker's or lab's recommendation, then adjust it by the gage's own history. Never use an overdue gage. And when a gage comes back out of tolerance, the job is not done when it is adjusted: look back at every part it measured since its last good calibration.

Why it matters on the floor

Every accept or reject decision trusts a gage. If the gage reads wrong, good parts get scrapped and bad parts get shipped, and nobody knows which. Calibration is how you know the number on the screen is the real number. The register is how you know it was true last month too.

What goes in the register

ColumnWhy it is there
Gage IDOne unique number, marked on the gage itself, so records point to one tool.
Description, range and resolutionSo you know it is the right gage for the tolerance it checks.
Location and ownerWhere it lives and who answers for it. Gages nobody owns go missing.
Interval, last date and next dueThe heart of it. Next due should work itself out from the last date and the interval.
Who calibrates itIn-house or the outside lab, and the method used.
Certificate numberSo an auditor, or you, can find the proof in a minute.
As-found resultPass or out of tolerance before adjustment. This is what tells you whether past parts are in question.
StatusIn use, due soon, overdue, out of service. Overdue gages should jump out at you.

How to set intervals

  • Start from the recommendation. The maker or your calibration lab will suggest an interval. Use it as the first guess, not the final answer.
  • Let the history move it. A gage that comes back out of tolerance gets a shorter interval. One that passes several times in a row can be stretched, carefully, with the results written down as the reason.
  • Weigh the job it does. Gages on tight tolerances, special characteristics or heavy daily use earn shorter intervals than a tape measure in maintenance.
  • Count use, not just days. A gage used on every part of every shift wears faster than the same model used once a week.

Labels, storage and the overdue gage

Every gage carries a label with its ID and next due date where the user sees it. Overdue gages come off the floor and get a do not use tag, not a promise to send them out next week. Store gages so they cannot be dropped, crushed or mixed up, and check the label is still readable on your layered process audits.

When a gage comes back out of tolerance

This is where most calibration programs fall short. The gage gets adjusted, a new sticker goes on, and nobody asks what it measured while it was reading wrong. That question is the whole point.

  1. Tag it and stop using it

    Pull the gage from use and tag it. Record the as-found readings from the calibration certificate before anyone adjusts or repairs it.

  2. Work out how far off it was

    Compare the as-found error with the tolerances it was used to check. A gage that was off by a tenth of the tolerance is a different problem from one off by half of it.

  3. Find what it measured

    Go back to the last calibration it passed. List the part numbers, lots and dates it checked since then, from inspection records, check sheets and the control plan.

  4. Decide if any product is at risk

    For each lot, ask: could a bad part have passed because of this error? Readings close to the limits are the ones at risk. Write the decision and the reasoning down.

  5. Contain what could be bad

    Re-check or hold suspect product in stock, in work in process and in transit. If suspect parts may have shipped, follow your customer's requirements for telling them.

  6. Fix the cause and the interval

    Ask why it drifted: wear, damage, the wrong gage for the job, or an interval too long. Shorten the interval if the history says so, and record the whole review in the register.

A worked example (made up)

A digital caliper, gage C-114, comes back from its six month calibration reading 0.05 mm high as found. Since its last good calibration it checked hole spacing on three part numbers, all with a tolerance of plus or minus 0.15 mm, on the final inspection check sheet.

The quality engineer pulls the check sheets for those six months. A 0.05 mm error matters only for readings within 0.05 mm of a limit. Most readings sat near nominal. Two lots had readings within 0.05 mm of the upper limit. A reading near the upper limit on a gage that reads high means the real part was smaller than recorded, so still inside tolerance; but readings near the lower limit could hide a part that was really undersize. One lot had three readings near the lower limit. That lot is still in finished goods, so it gets re-measured with a calibrated gage. All three parts pass.

The review goes in the register with the as-found data, the lots checked and the decision. The caliper's interval drops from six months to three, because it had been dropped twice, and the next two results will decide if three months is right.

Calibration is not the whole story

A calibrated gage can still be the wrong gage, or too variable in real hands on real parts. Calibration proves the gage reads true against a standard. Gage R&R proves the whole measurement system can tell good parts from bad. Before you blame the process for a drift, rule out the gage.

Mistakes that cause trouble

  • Gages that are not in the register. Personal tools, spare gages in a drawer, the fixture nobody thought of as a gage.
  • Reading the sticker, not the certificate. The as-found result is on the certificate. If nobody reads it, out of tolerance gages never trigger a look back.
  • A register nobody opens. Due dates only work if someone looks at them every week.
  • Stretching intervals without data. Longer intervals save money until the first out of tolerance result puts six months of parts in question.

Questions people ask

How often should gages be calibrated?

There is no single answer. Start from the maker's or calibration lab's recommendation, then adjust by history: shorten the interval when a gage comes back out of tolerance, and only extend it after several clean results in a row. Heavy use, rough handling and tight tolerances all argue for shorter intervals.

What is the difference between as-found and as-left?

As-found is how the gage read when it arrived for calibration, before any adjustment. As-left is how it reads after. The as-found result is the one that tells you whether past measurements can be trusted.

What do I do with an overdue gage?

Take it out of use and tag it until it is calibrated. If it was used while overdue, treat the measurements it made the same way you would treat an out of tolerance result: check them.

Is calibration the same as gage R&R?

No. Calibration checks a gage against a reference standard, so it tells you if the gage reads true. Gage R&R checks how much the whole measurement system varies, with real parts and real operators. You need both.

Can we calibrate gages in-house?

Many plants calibrate simple gages in-house against reference standards that are themselves calibrated by an outside lab, with a written method and trained people. Follow the requirement your customer or auditor cites for what must go to an accredited lab.

Do personal gages need to be in the system?

If a personal caliper or micrometer is used to accept parts, it needs an ID and a due date like any company gage. If it is not in the register, it should not be at the bench.